How to operate a rotorless vulcanizer

How to operate a rotorless vulcanizer

The rotorless vulcanizer has now become an indispensable instrument for the rubber industry. What should he do in the end?

1. Open the red button behind the vulcanometer, start the vulcanometer and turn on the computer and air compressor (air compressor is normally open)

2. In accordance with the experimental requirements, set the experimental temperature on the control panel of the vulcanizer and press ENTER to confirm.

3. Open the operation program of the rotorless vulcanizer on the desktop of the computer. Set the running time and torque range on the “parameter setting” of the program, and the data is automatically saved as “txt” file.

4. Create a new folder in the E drive. The “Storage File” in the program settings is the address where the data is saved and the data is placed in its own folder.

5. After the temperature reaches the experimental temperature, place a 10cm x 10cm cellophane on the mold cavity, place the glue on the glass paper, and place a piece of cellophane on the glue. Then click the "Start" button to start the experiment.

6. Save the data after the experiment and clean the mold cavity.

Note: No.1 and No.2 vulcanizers have a rotor vulcanizer, No. 3 vulcanizers have a rotorless vulcanometer, and a rotor vulcanometer has a large torque, which is suitable for rubbers with a large viscosity, and for rubbers with a low viscosity, such as Silicone rubber, liquid rubber, etc. are suitable for the No. 3 rotorless vulcanizer with less torque.